Title: Method for express determination of medical radionuclides 99mTc,67Cu concentration using spectrometer based on Si Planar detector
Authors: Deiev, Oleksey
Kiprich, S.
Maslov, N.
Ovchinnik, V.
Shulika, M.
Vasiliev, G.
Yalovenko, V.
Keywords: X-ray
pharmaceutical preparation
relative intensities
Gamma spectrum
Issue Date: 12-Dec-2017
Citation: Method for express determination of medical radionuclides 99mTc,67Cu concentration using spectrometer based on Si Planar detector / O. S. Deiev, S. K. Kiprich, N. I. Maslov, V. D. Ovchinnik, M. Y. Shulika, G. P. Vasiliev, V. I. Yalovenko // Problems of Atomic Science and Technology. Series “Nuclear Physics Investigations”. – 2017. – № 6 (112). – Р. 117–121.
Abstract: The method of express analysis provides the measurement of radiation spectra of medical radionuclides by a detection module based on an uncooled silicon planar detector and a "scintillator-silicon photosensor" type module. A spectrometric device for express analysis of radionuclide concentration has been developed and its prototyping has been carried out, confirming the operability of the device. The spectrometer analyzer is powered from the USB port of the PC (including a laptop) that provides the possibility of autonomous operation. Test measurements were carried out using the prototype of the spectrometer and X-ray sources, as well as spectral distributions of the radiation of real samples of 99mTc in a special container (glass ampoule). For a Si-PIN-detector with a thickness of 300 μm, the emission spectrum consists of a line of 99mTc with an energy of 140.5 keV and two peaks of the characteristic X-ray radiation (CXR) of technetium. The calculation in GEANT4 shows the matching of the 99mTc activities obtained from the registration of CXR and the main line. A measurement of the technetium concentration both by the 140.5 keV line and on the CXR radiation was proposed that significantly increases the speed of the data accumulation.
URI: https://repo.knmu.edu.ua/handle/123456789/18971
ISSN: 1562-6016
Appears in Collections:Наукові праці. Кафедра медичної та біологічної фізики і медичної інформатики

Files in This Item:
File Description SizeFormat 
117-121_7.d2_cor.doc677 kBMicrosoft WordView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.